Envent Engineering Ltd.
Located in Calgary, Canada, Envent Engineering is a world class developer of measurement devices for the natural gas processing industry. We produce H2S / Total Sulfur analyzers, Moisture monitors, O2 analyzers, CO2 analyzers, Gas Chromatographs and Low pressure flare vent / meters.
Recent News Release
Coming soon.
Head Office
Bay E, 7060 Farrell Road SE
Calgary, Alberta T2H 0T2
Canada (+ see map)
Frequently Asked Questions:
What cal-gas do I need?
How often should I calibrate the H2S analyzer?
How often change the samle system filter?
How much is the analyzer venting?
What should the sweep valve be set at?
How long will the tape last?
Does the Bypass time out?
How often does the humidifier need to be filled?
What are the analog output settings?
What is the range of the H2S analyzer?
Display is indicating a sensor fault.
What is the gain of the analyzer supposed to be?
Erratic H2S readings.
The analyzer was flooded, how do I fix it?
Analyzer reading incorrectly.
- What cal-gas do I need?
- How often should I calibrate the H2S analyzer?
- How often should I change the sample system filter?
- How much is the analyzer venting?
- What should the sweep valve be set at?
- How long will tape the tape last?
- Does the Bypass time out?
- How often does the Humidifier need to be filled.
- What are the Analog Output settings?
- What is the range of the H2S Analyzer?
- Display is indicating a Sensor Fault
- Put the Analyzer in Bypass
- Turn off sample gas at flowmeter
- Press the Menu/Set button until MTR RUN displayed
- Press Left or Right on the Display to move the tape to a new white piece of tape
- Remove the cover the sensor block. Press the small pushbutton located on the lower left side next to the wire connector. The sensor block will enter a "re-zero" procedure indicated by a red led.
- When the "re-zero"procedure is complete the green led should illuminate.
- Initiate another motor run
- Press the Menu/Set button until zero is displayed. Value should be between 900 & 1100 mV.
- Turn on sample gas to flow of two
- Take analyzer out of Bypass.
- What is the gain of the analyzer supposed to be?
- Erratic H2s readings
- The analyzer was flooded, how do I fix it?
- Analyzer reading incorrectly
Ideal cal-gas concentration would at the value normaly operating at or alarming at. Alternatively cal-gas to use is 3/4 concentration of the range of the analyzer.
Calibration checks should be done ever 3 to 6 months. Or as required by site preventive maintenance schedule.
The Filter should be changed as required. The frequency is dependant on the liquid content and particulates in the gas. A higher volume of a liquids and particulates will require more frequent changes. Filters should be checked when analyzer is calibrated.
The standard H2S analyzer vents 4 standard cubic feet per hour.
The sweep valve should be just opened slightly until flow can be felt from the tubing. This will reduce lag time and allow liquids to be removed.
The analysis time is dynamic and changes with the concentration of the H2S during the sample.
| Aperture | H2S (ppm) | Tape Life (Days)* |
|---|---|---|
| None | 4.0 ppm | 34.7 |
| 1/16th | 16.0 ppm | 30.4 |
*Tape life based on constant concentration sample over a averaged analysis time with a 300 ft tape.
By default the H2S analyzer does NOT time out when placed in bypass. A time out feature is available in the software, as well as Bypass Disable.
As long as there is some fluid in the humidifier it will be able to maintain the humidity of the gas. The fluid should be topped up each time the analyzer is calibrated or the tape is changed.
4-20mA wiring
The analog output of the H2S analyzer is a 4-20 mA signal calibrated at factory. The connections requires loop power from the PLC. If loop power is not able to be supplied, a powered AO board addition can be purchased.
| Symptom | Solutions |
|---|---|
| PLC Indicating 0 mA | Wiring to analyzer is incorrect. |
| Power to loop is not present. If loop power not supplied then Powered AO board option may be required. | |
| PLC Indicating ~2 mA | Analyzer in Bypass |
| PLC Indicating higher or lower signal than analyzer display | Scaling of analyzer is different setting on PLC |
Determining the range
The range of the analyzer is set at the time of purchase. The accuracy and tape life are optimized to the customers request. The optimized range can be determined by the calibration sheet found in the back of the manual.
The Envent H2S analyzer is capable of indicating 10x the optimized range of H2S concentration. Accuracy of the reading can not be guaranteed when indicating outside the optimized range.
Optimizing the range
| Range (ppm) | Aperture |
|---|---|
| 0-10 | None |
| 0-20 * | 1/16th |
| 0-50 | 1/32nd |
| 0-100 | Consult Envent Eng. |
For dilution analyzer systems the range adjusment depends on the dilution tubing. Please consult Envent about dilution range adjustments
Sensor Fault
The sensor Fault is a result of the sensor mV level being below 500 mV or above 1200 mV. Most often caused by an incorrect zero being achieved.
| Aperture | Gain* |
|---|---|
| None | 2.5-3 |
| 1/16th | 7.5-8.5 |
| 1/32nd | 10-13 |
| Possible Reasons | Solutions |
|---|---|
| Trigger slide not seated properly | Ensure Trigger slid is seated in the groove of the sample chamber |
| Sample vent blocked or insufficient. | Vent should be 1/2" tubing on a downward slope. Ending at shepards hook. Possible heat trace required on tubing |
| Contaminants in sample system | Sample system has been flooded or contaminated. Sample system requires cleaning. Refer to cleaning procedures in manual. |
Cleaning the analyzer
Once the analyzer has been flooded with contaminants it requires cleaning. Please refer to the Cleaning Procedure
After the cleaning is complete it is advisable to re-calibrate the analyzer.
| Symptom | Reasons | Solution |
|---|---|---|
| Analyzer reading higher H2S than expected | Gain too high. | Calibrate with cal-standard and adjust gain. |
| Vent line blocked | Check vent line for moisture build up and possible frozen sections. Review recommending venting. | |
| Aperture stip become dislodged | Reattach aperture strip with RTV 102 silicone | |
| Analyzer reading lower H2S than expected | Gain too low. | Calibrate with cal-standard and adjust gain. |
| Sample system contaminated | Refer to cleaning procedure to remove contaminants. | |
| Trigger slide not seated properly | Ensure trigger slide is firmly seated in groove of chamber block |