Model M90XP

Carbon Dioxide Monitor

Dual-Beam NDIR Technology

The Model M90XP and M90XP-ppm are designed to monitor ppm and percent level CO2 concentrations in natural gas and biogas applications. An infrared, dual-beam technology is incorporated into the design of the instrument. The M90XP and M90XP-ppm can be included as an additional measurement with other Envent analyzers, or can be offered as a stand-alone analyzer with an integrated sample conditioning system.

 

  • Features
    • Class I, Division 1
    • Dual-Beam NDIR Technology.
    • Long-term stability.
    • Low power consumption (2 watts).
    • Dual alarm outputs.
    • 4-20 mA output.
    • RS-232 and RS-485 Modbus communications.
    • Two Adjustable set point control relays.
    • Ppm and % level sensors available. 
  • Downloads
  • Certification
  • Applications
    • CO2 is measured routinely in natural gas for product quality and process control purposes. Common ranges are from 0-5% for sales gas and up to 0-30% for inlet CO2 concentration.
    • Gas plant Inlet.
    • Certain H2S-removal chemical processes can be fine-tuned by CO2 slipstreaming.
    • Quantify CO2 as a combustion by-product.
    • Biogas applications
  • Specifications
    Electrical classification Class I, Div1 Groups B, C & D
    Operating Principle Dual-beam, non-dispersive infrared (NDIR).
    Detector Sapphire window with glass frit seal
    Sensing Cell Proof Pressure (Sample cell only) 2 PSI (13.8 KPa) gauge pressure
    Measurement Range 0-5000ppm; 0–20% or 0-100% CO2; Custom ranges available
    Repeatability ± 0.1 % CO2
    Drift Less than 2.5% of reading/year
    Accuracy ± 3% of reading or 0.1% CO2 at cal temperature,
    ± 5% of reading or 0.1% CO2 over full operating temperature range.
    Warm U Time Less than 5 minutes
    Operating Humidity Sensing Cell: 0–100% RH (non-condensing).
    Electronics: 0 – 90% RH (non-condensing).
    Storage Temperature -20 to 60°C (-4 to 140°F).
    Storage Humidity 0–90% RH (non condensing).
    Analog Output Isolated 4–20 mA
    Digital Interface RS-485 & RS-232
    Operating Power 24 VDC at 2 Watts standard.
    12 VDC, 120/240 VAC optional
    Source Life 10 years minimum
    Alarm Outputs 2 solid state drivers 2 Amp, 30 VDC maximum
    Standard Sample System Inlet filter (1500psig/103bar max).
    SS single stage regulator (3000psig/207bar max).
    Flow meter (100psig/6.9bar max).

Contact us to learn more about this product

M-Series Monitors

M70XP Moisture Monitor

The Model M70XP is designed to monitor moisture/water content in natural gas and biogas applications.

M90XP Carbon Dioxide Monitor

The Model M90XP is designed to monitor percent level CO2 concentrations in natural gas and biogas applications.

M80XP Methane Monitor

The Model M80XP is designed to monitor CH4 in natural gas and biogas applications.

M60XP Oxygen Monitor

The Model M60XP is designed to monitor PPM and percent level O2 concentrations in natural gas and biogas applications.